Electrostatic Voltmeter - Isoprobe® - model 279
multi-channel applications
• Full complement of interchangeable
probes
• Selectable recorder output range
• LED meter with “black out” for dark
room applications
Side- and end-viewing probes accommodate various mounting requirements. Specialty high-resolution probes allow the 279 to resolve spot diameters as small as 1 mm. Probes include air purge capability to prevent ingress of particulate.
Range: ±3000
volts, auto polarity
Accuracy:* 0.1%,
±0.003%/°C over +20 to +40°C range (at recorder output) Useable to +50°C
Speed of Response:*
For typ. input step: <3ms (10% to 90%)
for 1kV step: <2½ms (10% to 90%)
for 2kV step: <3½ms (10% to 90%)
for 3kV step: <4ms (10% to 90%)
Settling time: <3½ms to 1% of 1kV step input
Frequency response: Small signal frequency response typical ±3db to >300Hz
Drift:* <0.01V/hr
after 1 hr warm-up (0.003V/hr typical). Not measureably affected by 10°C
temperature variation or changes between 10% and 90% relative humidity.
Noise:* <0.3V
rms or 2 volts peak-to-peak wide band (bandwidth restricted to <1kHz)
referred to input.
Surface resolution: Determined
by probe aperture size and surface-probe separation. Standard type 1034E &
S probes with 0.07” (1.75mm) aperture will resolve a 0.10” (2.5mm) spot at
0.02” (0.5mm) separation.
Recorder output: Compressed
analog output is input divided by 100, 200, or 1000 (factory selectable).
Typical maximum dynamic range at isolated BNC connector is > ±18 volts for
external loads >10kΩ
Output filter: Bessel low pass filter with 0.7ms constant delay
Probe: Uses
Model 1034 series probes
Power requirement: 100, 115, 230 VAC, ±10%, 50/60Hz, 15 watts
Size: 1.7x8.2x15.1”
(44x208x384mm), 1.75” rack mounting available (1 or 2 per rack)
Weight: 4lb.
(1.8kg)
Accessories: Manual, probe hole
vent plug, line cord, optional 220V fuses
CE Compliance: Pending
- Electrophotographic measurements
- Radiation effect studies
- Electret research
- Electric field studies
- Process monitoring and control
- Integrated circuit manufacturing and handling