Trek PD06078 four (4) channel electrostatic voltmeter
•Measurement Range for Each Channel:
±10 kV DC or peak AC
•Voltage Output Monitors and 4-20 mA Current Loop outputs provide accurate and concise measurement results for each ESVM channel
•Front Panel LED provides visual monitoring for each channel (switch selectable)
•Two Types of Probes Available:
A Side View Probe and a 45° Angle probe
•Chopper probes are DC stable with or without incident air ion flow
•Drift-Free Measurements
Trek Model PD06078 Electrostatic Voltmeter (ESVM) is a four (4) channel voltage measuring instrument which is ideally suited to monitor critical operations associated with semiconductor, LCD, electronic assembly, and other processes where static charge accumulation poses a threat to production yields or product quality.
The Model PD06078 contains four (4) Electrostatic Voltmeters in a single enclosure. Each channel features a measurement range of 0 to ±10 kV and accuracy better than 5% reading plus ±0.2% of full scale over the probe to surface separation distance of 15 to 30 mm.
Dimensions
177 mm H x 203 mm W x 228 mm D
(7” H x 8" W x 9" D).
Weight
8 lbs. (3.6 kg)
Power Supply Options
A two (2) pin [Mate-n-Lock] connector
or a DC power input for 24 V DC ±10%, 1.5 A supply, 3 A initial peak current.
(An Universal Power Adapeter is optional.)
Output Connector
(one for each channel)
A six position output connector
(6P/4C) provides connections for
the output voltage monitor, the
current output, and a ground
connection.
Operating Conditions Temperature
15 °C to 35 °C.
Relative Humidity
5% to 85%, noncondensing.
Power ON/OFF
Front panel switch.
Probe Connector Locations
Rear panel.
Four (4) Position Channel Switch
Selects one of the four channels outputs for display on the DPM.
Digital Panel Meter (DPM)
3½ digit LED voltage display.
Range
0 to ±10.00 kV.
Resolution
10 V.
Zero Offset
Less than or equal to ±2 counts.
ZERO Control
(one for each channel)
Rear panel potentiometer used to
produce zero volts output when
probe is coupled to a known zero
voltage source.
Monitor Output
(one for each channel)
Scale Factor
The voltage monitor is scaled at
1/1000th of the measured
voltage for each channel.
Output Noise
Less than 10 mV rms (using the side view probe and measured using the true rms feature of the Hewlett Packard Model 34401A digital voltmeter).
Output Impedance
47 Ω.
Current Output
(one for each channel)
Provides a current of 4-20 mA that
represents measured voltages
-10 kV to +10 kV.
Ground Connections
Four (4) banana jacks and a
designated terminal on the output
connector are all tied to chassis
ground.
The probes are chopper stabilized for drift-free operation.
Model 542P-45D
45° Orientation
Aperture size of 3.8 mm (0.15") diameter.
Dimensions
11 mm dia. x 79 mm L
(0.43" dia. x 3.1" L).
Model 542P-S
Side Orientation
Aperture size of 4 mm (0.156") diameter.
Dimensions
13 mm dia. x 75 mm L
(0.51" dia. x 3" L).
Probe Cable Length
5 meters (16 ft), nominal
[up to 10 meters (32.8 ft)
optional on side-view probe].
Probe-to-Surface Separation
15 mm to 30 mm (recommended).
(Vacuum probes are available)